MOCOM Autoclave Error Codes
E000 POWER OUTAGE
E001 OVERVOLTAGE
E002 INSUFFICIENT WATER QUALITY
E003 POOR QUALITY WATER
E004 LINE FREQUENCY ERROR
E007 FAN PROBLEM
E008 WATER CONDUCTIVITY ERROR
E009 WATER CONDUCTIVITY ERROR
E010 DOOR OPEN
E020 DOOR LOCK ERROR
E021 DOOR LOCK ERROR
E022 DOOR LOCK ERROR
E030 WATER LEVEL ERROR
E031 WATER LEVEL ERROR
E042 WATER LEVEL ERROR
E050 RUN VACUUM TEST
E051 RUN HELIX TEST
E052 RUN VACUUM+HELIX TEST
E060 ETHERNET CONFIGURATION ERROR
E061 WI-FI CONFIGURATION ERROR
E070 CLOSE THE DOOR
E126 SOFTWARE UPDATE IN PROGRESS
E130 OFFLINE DISABLED
E141 CLOUD SOFTWARE VERSION ERROR
E900 VACUUM TEST FAILED
E901 VACUUM TEST FAILED
E902 VACUUM TEST FAILED
E998 REMOTE SERVICE ACTIVE
E999 MANUAL INTERRUPTION
A032 LEVEL SENSOR PROBLEM
A040 WATER INLET FAILURE
A042 WATER FILLING MAXIMUM
A101 PT1 BROKEN (STERILIZATION CHAMBER)
A102 PT2 BROKEN (STEAM GENERATOR)
A103 PT3 BROKEN (HEATING ELEMENT)
A105 PT5 BROKEN
A111 PT1 SHORT CIRCUIT (STERILIZATION CHAMBER)
A112 PT2 SHORT CIRCUIT (STEAM GENERATOR)
A113 PT3 SHORT CIRCUIT (HEATING ELEMENT)
A115 PT5 SHORT CIRCUIT
A116 ADC ERROR ON PCB
A117 DOOR MOTOR OVERCURRENT
A120 OPERATING BOARD ERROR 3
A121 PROCESSOR CARD ERROR
A122 PROCESSOR CARD ERROR
A126 WI-FI MODULE ERROR
A131 SOLENOID VALVE 1 FAULTY
A132 SOLENOID VALVE 2 FAULT
A133 SOLENOID VALVE 3 FAULT
A134 SOLENOID VALVE 4 FAULT
A135 SOLENOID VALVE 5 FAULT
A136 SOLENOID VALVE 6 FAULT
A140 CLOUD FW UPDATE ERROR
A145 FAULTY CURRENT DRAWING
A146 SOLENOID VALVE DRIVER ERROR
A147 DOOR MOTOR DRIVER ERROR
A201 STEAM GENERATOR RESISTANCE OPEN CIRCUIT
A202 HEATING BAND OPEN CIRCUIT
A250 1PV TIMEOUT
A251 ATM1 UPTAKE
A252 1PP UPTAKE TIME TIMEOUT
A253 2PV TIMEOUT
A254 ATM2 UPPULSE TIMEOUT
A255 2PP UPPULSE TIMEOUT
A256 3PV TIMEOUT
A257 ATM3 UPPULSE TIMEOUT
A258 3PP UPPULSE TIMEOUT
A260 ATM3 DOWNPULSE TIMEOUT
A261 PRESSURE LEVELING TIMEOUT
A262 PRESSURE UPPULSE TIMEOUT
A353 ATM1 DOWNPULSE TIMEOUT
A356 ATM2 DOWNPULSE TIMEOUT
A360 SPD PRESSURE STAGE
A362 PRESSURE DOWNPULSE TIMEOUT
H150 PRESSURE SENSOR OPEN CIRCUIT
H160 SENSOR SHORT CIRCUIT
H400 PCONV/T RATIO UNBALANCED
H401 T/P CONVERSION RATIO UNBALANCED
H402 TEMPERATURE ABOVE MAX LIMIT
H403 TEMPERATURE BELOW MIN LIMIT
H404 TEMPERATURE VARYING ABOVE LIMIT
H405 PRESSURE ABOVE MAX LIMIT
H406 PRESSURE BELOW MINIMUM LIMIT (STERILIZATION PHASE)
H410 TIME MEASUREMENT ERROR
H411 STERILIZATION TIME ERROR
H990 OVERPRESSURE
H991 OVERHEATING (STERILIZATION CHAMBER, PT1)
H993 OVERHEATING (LAYER RESISTANCE, PT3)
For your autoclave technical service, new autoclave sales, used autoclave sales needs, as well as autoclave troubleshooting, maintenance, repair, installation, and remote assistance requirements, please feel free to contact us 24/7.