You can request 24/7 technical support by calling +902165195853 or contacting us via WhatsApp.

No matter where you are in the world, we provide remote support for Panoramic and Tomographic X-ray software malfunctions.

We provide remote and on-site technical support worldwide.

acteon x-ray

ACTEON PANORAMIC X-RAY ERROR CODES

0 / 1 Internal MCU error

500 ÷ 505 MCU Ethernet errors

100 / 101 Configuration area parameter doesn't match the expected one

102 Wrong version number in configuration area

103 / 104 Timeout error occurred during an EEPROM erase/write operation

200 Zero position optical sensor of rotation axis always activated

201 Zero position optical sensor never activated

202 / 203 Zero position optical sensor of rotation still active after exiting from zero

204 Unexpected activation of rotation optical sensor

205 Timeout on rotation

240 Zero position micro Y always active

241 Zero position micro Y never active

243 Timeout on Y axes

270 / 271 Hardware key fault

360 RX button pressed on start-up or before exam

362 RX button released during emission

370 Sensor ready lost during exposure

371 Sensor not ready

374 The computer connection drops or times out during exam

375 Sensor took long in configuration mode (while in preheat)

750 Generator board initialization error

751 Alarm "overvoltage kV"

752 Alarm "overload on filament" on Generator board

753 Alarm "overload anodic current"

754 Alarm "filament not OK"

755 Alarm "backup timer"

756 Alarm "PFC not OK"

757 Alarm "Brown OUT"

758 Alarm "NO X-ray"

759 Alarm "unexpected emission"

760 Alarm "NO RX button command"

761 Alarm "NO X-ray emission"

762 Bad unit status: emission flag detected unexpectedly

763 kV analog feedback out of range

764 mA analog feedback out of range

765 Filament analog feedback out of range

766 Generator board reset due to a brown out

767 Generator board reset due to low voltage detection

768 Generator board reset due to a watchdog timeout

769 Generator board reset due to a stack overflow

770 Mismatch between generator board (A2) and MCU board (A1) types (2D / 3D)

850 One or more keycodes are pressed

852 Button >O< pressed during movements

1200 DLL communication error

1201 Setup menu: write data EEPROM failure

1202 Unexpected value detected by the software

1203 Software allocation failure

1204 Exposure parameters failure

1205 Image buffer allocation failure

1401 Sensor frame lost during exam

1402 Sensor configuration failure

1403 Software watchdog error

1404 Sensor does not detect X-rays during exam

1405 Sensor frame lost during exam

1406 Error in sensor frame rate

X-ray technical service, new and used X-ray sales needs, as well as X-ray troubleshooting, maintenance, repair, installation, and remote assistance requirements, please feel free to contact us 24/7.

Benk Instruments Dental, Medical and Laboratory Equipments Technical Services
  • LinkedIn

©2026, Gökhan Benk

SHOPPING
Distance Selling Agreement
Privacy and Security
Cancellation and Return Policy
Personal Data Policy

visa_mastercard.png
bottom of page