FONA X-Ray Technical Service
Dental Panoramic Cephalometric Tomographic X-ray
You can request 24/7 technical support by calling +902165195853 or contacting us via WhatsApp.
No matter where you are in the world, we provide remote support for Panoramic and Tomographic X-ray software malfunctions.
We provide remote and on-site technical support worldwide.

FONA PANORAMIC X-RAY ERROR CODES
A X-ray generator hot
B CAM init
C Adjust Arm Position
D Wait PC READY
1 kV reference signal out of range
2 mA reference signal out of range
3 Carriage rotation encoder failure
4 Carriage displacement encoder failure
5 Carriage rotation sensor failure
6 Carriage displacement sensor failure
7 Collimator displacement sensor failure
10 Tube-head temperature exceeding limit Wait for tube cool down
11 Exposure aborted during irradiation
12 Exposure aborted before irradiation
13 38 V DC supply out of range
14 24 V DC supply out of range
15 15 V DC supply out of range
16 5 V DC supply out of range
17 High voltage failure
19 Exposure time exceeded
20 Exposure aborted after irradiation
21 No tube current
23 Filament current out of range
25 Thermal Sensor Failure
26 Battery fault (low voltage)
28 Collimator encoder failure
30 PC-Control unit communication failure during exam
32 Control unit fault
X-ray technical service, new and used X-ray sales needs, as well as X-ray troubleshooting, maintenance, repair, installation, and remote assistance requirements, please feel free to contact us 24/7.