VATECH PANORAMIC X-RAY ERROR CODES

001 The tube is not ready for use

002 The cable between the tube tank and Inverter board are disconnected

003 A current of the inverter board exceeds the maximum allowable level during X-ray irradiation

004 There is ±10kV or more voltage difference in tube voltage compared to a reference value

005 There is ±0.5mA or more current difference in tube current compared to a reference value

006 There is ±20kV or more voltage difference in tube voltage feedback compared to the average value

007 There is ±1mA or more current difference in tube current feedback compared to the average value

008 The temperature of the mono tank is above the setting temperature

009 The inverter output current is higher than 1A during X-ray irradiation (In EP, IP condition)

010 The inverter board falsely recognizes the exposure switch signal as OFF after the irradiation On command

011 The X-ray OFF command is not sent to the inverter board in 0.5 seconds after turning off the exposure switch

012 kV feedback is over -20kV compared to the setting value during X-ray irradiation

013 kV feedback is over +20kV compared to the setting value during X-ray irradiation.

014 The mA feedback value is less than 50% compared to setting conditions during X-ray irradiation.

015 The mA feedback value is higher than 150% compared to setting conditions during X-ray irradiation.

021 Rotator-axis motor origin movement

027 Cephalo sensor motor origin movement

037 Generator tilting

060 The exposure switch is pressed when turning on the equipment.

061 X-ray exposure is stopped by releasing the exposure switch.

102 There is no response during CAN communication.

X-ray technical service, new and used X-ray sales needs, as well as X-ray troubleshooting, maintenance, repair, installation, and remote assistance requirements, please feel free to contact us 24/7.

Benk Instruments Dental, Medical and Laboratory Equipments Technical Services
  • LinkedIn

©2026, Gökhan Benk

SHOPPING
Distance Selling Agreement
Privacy and Security
Cancellation and Return Policy
Personal Data Policy

visa_mastercard.png
bottom of page